Power Requirements: Two "AA" alkaline batteries or high capacity rechargeable batteries (NimH)
Dimensions: 124 x 70 x 33 mm (approximate)
Weight: 270g (including batteries, approximate)
Operating Ambients: Temperature: 10°C (50°F) to 50°C (122°F)
Humidity: 20% to 95% RH non-condensing
Storage Ambients: Temperature: -20°C (-68°F) to 70°C (158°F)
Humidity: 15% to 95% RH non-condensing
Altitude: -170 to 1,700 meters
Gauging Ranges: Systolic Blood Pressure: 60 to 250 mmHg
Diastolic Blood Pressure: 25 to 200 mmHg
Maximum Inflation: 270 mmHg
Heart Rate: 40 to 200 bpm
Memory: Up to 250 readings using Alkaline batteries
Up to 110 reading using rechargeable NimH
Method of Measurement: Oscillometric with step deflation
Accuracy: ± 3 mmHg
International Standards:
· EN 60601-1-1:2001 ; Medical electrical equipment - Part 1-1: General requirements for safety. Collateral standard: Safety requirements for medical electrical systems.Specific requirements for the protection against hazards of ignition of flammable anesthetic mixtures and to control the risk of fire (Fire prevention)
· EN 60601-2-30:2000 ; Medical electrical equipment - Part 2-30: Particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressuring monitoring equipment
· EN 60601-1-2:2001 ; Medical Electrical Equipment - Part 1-2: General requirements for safety. Collateral Standard: Electromagnetic compatibility - requirements and tests
· EN 1060-1:1996 ; Specification for Non-invasive sphygmomanometers - Part 1.
General requirements
· EN 1060-3:1997 ; Non-invasive sphygmomanometers - Part 3: Supplementary
requirements for electro-mechanical blood pressure measuring systems
· AAMI SP 10 ES1 Category C: 1992 (battery powered) Electronic or automated sphygmomanometers
Calibration: Minimally, once per year
Safety System: Maximum Inflation pressure limited to 300 mmHg.
Auto safety release valve for power failure.
Maximum BP measurement time limited to less than 180 seconds
Data Connector: Stereo mini-plug headphone type connector
Operator Control: 1-button control and LCD
Automatic Measurement Intervals: Programmable up to four separate time periods at 5 to 120 minute intervals
|